Improving radiofrequency power and specific absorption rate management with bumped transmit elements in ultra-high field MRI.
Alireza Sadeghi-TarakamehGregor AdrianyGregory J MetzgerRussell L LagoreSteve JungstLance DelaBarrePierre-François Van de MoorteleKamil UgurbilErgin AtalarYigitcan EryamanPublished in: Magnetic resonance in medicine (2020)
In this study, we investigated a simple design strategy based on adding bumps to a resonant element to reduce the local SAR while maintaining B 1 + inside an ROI. As an example, we modified an FD and performed EM simulations and phantom experiments with a 10.5T scanner. Results show that the peak 10g-averaged SAR can be reduced more than 25%.