Login / Signup

Towards an integrated neonatal brain and cardiac examination capability at 7 T: electromagnetic field simulations and early phantom experiments using an 8-channel dipole array.

Jérémie ClémentRaphaël Tomi-TricotShaihan J MalikAndrew WebbJoseph V HajnalÖzlem Ipek
Published in: Magma (New York, N.Y.) (2022)
The 8-channel dipole array design integrated for neonatal brain and cardiac MR was successfully demonstrated, in simulation with coverage of the baby and increased SAR efficiency levels compared to the birdcage. We conclude that the 8Tx-dipole array promises safe operating procedures for MR imaging of neonatal brain and heart at 7 T.
Keyphrases