Inferior load generated by preloaded versus manually loaded haemorrhoid banding devices: the effect of 'creep relaxation'.

Jayson MoloneyRichard HiscockM Hung Nguyen
Published in: ANZ journal of surgery (2019)
The load generated by haemorrhoid bands from preloaded devices is lower and deteriorates significantly towards their expiry date compared with bands from manually loaded devices. This is mostly likely due to their storage in a stretched state. This should be considered by clinicians when using haemorrhoid banding devices.