Login / Signup

Subcutaneous implantable cardioverter-defibrillator noise following left ventricular assist device implantation.

Banveet Kaur KhetarpalAwad JavaidJustin Z LeeFred KusumotoSiva K MulpuruDan SorajjaYong-Mei ChaKomandoor Srivathsan
Published in: Journal of arrhythmia (2023)
The incidence of LVAD-related S-ICD noise is high in patients with concomitant LVAD and S-ICD with significant impact on device function. As conservative management failed to resolve the EMI, the S-ICDs had to be programmed off to avoid inappropriate shocks. This study highlights the importance of awareness of LVAD-SICD device interference and the need to improve S-ICD detection algorithms to eliminate noise.
Keyphrases
  • left ventricular assist device
  • air pollution
  • machine learning
  • risk factors
  • deep learning
  • loop mediated isothermal amplification
  • drug induced
  • quantum dots