A comparison of static and dynamic ∆B0 mapping methods for correction of CEST MRI in the presence of temporal B0 field variations.

Esau Poblador RodriguezPhilipp MoserBarbara DymerskaSimon RobinsonBenjamin SchmittAndre van der KouweStephan GruberSiegfried TrattnigWolfgang Bogner
Published in: Magnetic resonance in medicine (2019)
Reliable correction of scanner instabilities is essential to establish clinical CEST MRI. The three dynamic approaches presented improved the ∆B0 -correction performance significantly in the presence of frequency drift compared to established static methods. Among them, the self-corrected CEST-GRE-2TE was the most accurate and straightforward to implement.