Login / Signup

Very complex internal standard response variation in LC-MS/MS bioanalysis: root cause analysis and impact assessment.

Nico C van de MerbelRemco A KosterCorey Ohnmacht
Published in: Bioanalysis (2019)
Internal standards (ISs) are essential for the development and use of reliable quantitative bioanalytical LC-MS/MS methods, because they correct for fluctuations in the analytical response that are caused by variations in experimental conditions. Sample-to-sample differences in the IS response are thus to be expected, but a large variability often is an indication of nonoptimal sample handling or analysis settings. This paper discusses a number of cases of very complex variation of IS responses that could be attributed to analytical problems such as injection errors and sample inhomogeneity, and matrix-related issues such as degradation and increased ionization efficiency. A decision tree is proposed to help find the underlying root cause for extreme IS variability.