Login / Signup

A Sample-In-Answer-Out Microfluidic System for the Molecular Diagnostics of 24 HPV Genotypes Using Palm-Sized Cartridge.

Rui WangJing WuXiaodong HePeng ZhouZuojun Shen
Published in: Micromachines (2021)
This paper proposes an automated microfluidic system for molecular diagnostics that integrates the functions of a traditional polymerase chain reaction (PCR) laboratory into a palm-sized microfluidic cartridge (CARD) made of polystyrene. The CARD integrates 4 independent microfluidic sample lanes, which can independently complete a sample test, and each sample lane integrates the 3 functional areas of the sample preparation area, PCR amplification area, and product analysis area. By using chemical cell lysis, magnetic silica bead-based DNA extraction, combined with multi-PCR-reverse dot hybridization with microarray, 24 HPV genotypes can be typing tested in CARD. With a custom-made automated CARD operating platform, the entire process can be automatically carried out, achieving sample-in-answer-out. The custom-made operation platform is developed based on a liquid handling station-type, which can automatically load off-chip reagents without placing reagents in CARD in advance. The platform can control six CARDs to work simultaneously, detect 24 samples at a time. The results show that the limit of detection of the microfluidic system is 200 copies/test, and the positive detection rate of clinical samples by this system is 100%, which is an effective method for detection of HPV.
Keyphrases