Login / Signup

Efficacy and safety testing of a COVID-19 era emergency ventilator in a healthy rabbit lung model.

Luke A WhiteBenjamin S MaxeyGiovanni F SolitroHidehiro TakeiSteven A ConradJ Steven Alexander
Published in: BMC biomedical engineering (2022)
Although conventional ventilators are always preferable outside of emergency use, the FALCON ventilator safely and effectively ventilated healthy rabbits without lung injury. Emergency ventilation using accessible and inexpensive strategies like the FALCON may be useful for communities with low access to medical resources and as a backup form of emergency ventilation.