Login / Signup

Home Blood Pressure Monitoring Devices: Device Performance in an Alaska Native and American Indian Population.

Krista R SchaeferAmber L Fyfe-JohnsonCarolyn J NoonanMichael R ToddJason G UmansDorothy M CastilleRobert RosenmanDedra S BuchwaldDenise A DillardRenee F RobinsonClemma J Muller
Published in: Journal of aging and health (2021)
Objectives: Home blood pressure monitoring (HBPM) is an important component of blood pressure (BP) management. We assessed performance of two HBPM devices among Alaska Native and American Indian people (ANAIs). Methods: We measured BP using Omron BP786 arm cuff, Omron BP654 wrist cuff, and Baum aneroid sphygmomanometer in 100 ANAIs. Performance was assessed with intraclass correlation, paired t-tests, and calibration models. Results: Compared to sphygmomanometer, average BP was higher for wrist cuff (systolic = 4.8 mmHg and diastolic = 3.6 mmHg) and varied for arm cuff (systolic = -1.5 mmHg and diastolic = 2.5 mmHg). Calibration increased performance from grade B to A for arm cuff and from D to B for wrist cuff. Calibration increased false negatives and decreased false positives. Discussion: The arm HBPM device is more accurate than the wrist cuff among ANAIs with hypertension. Most patients are willing to use the arm cuff when accuracy is discussed.
Keyphrases