Investigating the discrepancy between MAIA and MP-1 microperimetry results.

Lixing XuZhichao WuRobyn H GuymerAndrew J Anderson
Published in: Ophthalmic & physiological optics : the journal of the British College of Ophthalmic Opticians (Optometrists) (2021)
While MAIA measurements showed low bias compared with our rigorously determined reference standard, the MP-1 showed large discrepancies that could not be explained purely by the limited dynamic range of the instrument. MAIA and MP-1 sensitivity values cannot be compared directly, and caution is advised when assessing absolute sensitivities or eccentricity effects in the extensive MP-1 literature.