Validation of noninvasive blood pressure monitoring function of EDAN elite V5 patient monitor with reference invasive measurement according to the ISO 81060-2:2018 standard.

Yi WanDuan LiJin-Feng ChenYang Zhang
Published in: Blood pressure monitoring (2024)
The noninvasive blood pressure monitoring function of the EDAN elite V5 patient monitor passed all the requirements of ISO 81060-2:2018 and can be recommended for clinical use in adults, adolescents, or children.