Predictors of a high defibrillation threshold test during routine ICD implantation.
B Al-AtiaBert VandenberkG VörösC GarwegJ EctorRik WillemsPublished in: Acta cardiologica (2017)
A high DFT is still present in about 5% of patients and is more frequent in patients with a severely dilated left ventricle, a very low LVEF, right sided pre-pectoral implantation and wider QRS duration. It might be clinically important to continue DFT testing in these high risk patients.