Login / Signup

A systematic review and meta-analysis on oncological radiotherapy in patients with a cardiac implantable electronic device: Prevalence and predictors of device malfunction in 3121 patients.

Vincenzo Livio MalavasiJacopo Francesco ImbertiAlberto TosettiGiulio Francesco RomitiMarco VitoloMassimo ZecchinErcole MazzeoDe Marco GiuseppinaFrank LohrTeresa Lopez-FernandezGiuseppe Boriani
Published in: European journal of clinical investigation (2022)
Radiotherapy related CIEDs malfunction had a prevalence ranging from 4% to 20%. The use of neutron-producing energies and more complex devices (ICD/CRT-D) were associated with higher risk of device malfunction, while the radiation dose at CIED did not significantly impact on the risk unless higher doses (>10 Gy) were used.
Keyphrases