Performance evaluation of three i-SENS glucometers using arterial blood samples compared with the YSI 2300 Glucose Analyzer.

Ha Nui KimKyung Chul MoonSoo-Young Yoon
Published in: Journal of clinical laboratory analysis (2020)
The i-SENS glucometers showed stable and accurate results when leftover arterial blood samples were used. Therefore, POC glucometers could be useful in critical care settings, such as intensive care units, where arterial samples are routinely used.
Keyphrases
  • intensive care unit
  • high resolution
  • type diabetes
  • blood glucose
  • blood pressure
  • weight loss