Evaluation of acquisition modes for semi-quantitative analysis by targeted and untargeted mass spectrometry.
Hannah M BrittTristan CragnoliniSuniya KhatunAbubakar HatimyJuliette JamesNathanael PageJonathan P WilliamsChristopher HughesRichard DennyKonstantinos ThalassinosJohannes P C VissersPublished in: Rapid communications in mass spectrometry : RCM (2022)
A comparison of the acquisition modes revealed that data-independent acquisition and parallel reaction monitoring methods are most robust for semi-quantitation when considering achievable LDR and LLOQ. IM-enabled modes exhibited sensitivity increases, but a simultaneous reduction in dynamic range required correction methods to recover. These findings will assist users in identifying the optimum acquisition mode for their analyte quantitation needs, supporting a diverse range of applications and providing guidance for future acquisition mode developments.