Performance Evaluation of the DxC 700 AU Chemistry Analyzer in Hemoglobin A1c Measurement

Yu Jeong ChoiHyein KangChan-Ik ChoJohn Hoon RimSang Guk LeeJong Baeck Lim
Published in: Annals of laboratory medicine (2022)
The overall analytical performance of DxC 700 AU was acceptable. The device is accurate, precise, and linear over a wide HbA1c concentration range. Although DxC 700 AU results highly correlated with those of Cobas c 513, caution should be exercised in cases of high HbF and HbH concentrations.