Login / Signup

Rapid diagnostics of orthopedic implant-associated infections using Unyvero ITI implant and tissue infection application is not optimal for Staphylococcus species identification.

Hege Vangstein AamotBjørn Odd JohnsenInge Skråmm
Published in: BMC research notes (2019)
72 tissue biopsies from 15 consecutive patients with deep OIAI infections were analyzed with conventional culturing including phenotypic antibiotic susceptibility testing and the U-ITI. U-ITI showed lower sensitivity than conventional culturing concerning detection of pathogens (73% vs 93%). 4/15 patients would have been given false negative results by U-ITI, all of which were culture-positive for Staphylococcus species. Median time to detection of pathogens was 47 h and antibiotic resistance 89 h by conventional methods compared to 13.5 h with the U-ITI. The U-ITI did not detect antibiotic resistance, whereas conventional culturing showed resistance to antibiotics covered by the U-ITI panel in 2 patients. Time to detection of pathogens was improved, but the detection limit for staphylococci was unsatisfactory. Although the time to antibiotic treatment recommendation was significantly reduced, the U-ITI would have resulted in incorrect antibiotic recommendation in 2 patients. Our data do not support use of this assay in diagnostics.