Analytical validation of a new point-of-care assay for serum amyloid A in horses.

Diana SchwartzNicola PusterlaS JacobsenM M Christopher
Published in: Equine veterinary journal (2018)
The POC test has acceptable accuracy and precision in equine serum/plasma with SAA concentrations of up to at least 1000 μg/mL. Low inter-batch precision at high concentrations may affect serial measurements, and the use of the same test batch and sample type (serum/plasma or whole blood) is recommended. Comparison of results between the POC test and the TIA is not recommended.