Dissociation and Subsequent Stable Malrotation of Femoral Stem Within Sleeve After Sivash-Range of Motion Total Hip Arthroplasty: A Case Report.

Neil PathakErik McDonaldZachary J RadfordDavid H Gibson
Published in: JBJS case connector (2023)
We present a rare complication of S-ROM THA-dissociation of the femoral stem from the proximal sleeve, with subsequent re-engagement in static malrotation. No corrosion was observed, suggesting that the stem had stably reseated within the sleeve.