Tomosynthesis implementation with adaptive online calibration on clinical C-arm systems.
Khanlian ChungLothar R SchadFrank Gerrit ZöllnerPublished in: International journal of computer assisted radiology and surgery (2018)
The circular tomosynthesis was tested on a ball bearing and an anthropomorphic phantom. The results show that the calibration is robust against isocenter shifts and object deviations in the CBCT. All reconstructions used 100 projections and displayed limited angle artifacts. The accuracy of the positions and shapes of high-contrast objects were, however, determined precisely. (The maximal center position deviation is 0.31 mm.) CONCLUSION: For medical procedures that primarily determine the precise position of high-contrast objects, circular tomosynthesis could offer an approach to reduce dose exposure.