[Low-field magnetic resonance imaging : Just less expensive or completely different?]

Jürgen Hennig
Published in: Der Radiologe (2022)
Over the years the development of field strength in magnetic resonance imaging (MRI) has continued to increase from the low-field systems in the early years (0.2-0.5 T) to 1.5 T to 3 T to 7 T and more. In the last 2 years, there has been a renewed interest in MRI at lower fields, mainly driven by the development of "dry" superconductive magnets. The following article demonstrates that this renewed interest for lower fields is not a déjà vu purely driven by economic needs. The field strength appears to be from yesterday, but the combination with the tremendous improvements and innovations of all relevant components-gradients, radiofrequency system and especially new algorithms including the use of artificial intelligence (AI)-allow the realization of diagnostically adequate MRI without compromise in patient throughput and efficiency. In addition to the lower field, there are also some inherent advantages, e.g., for MRI of the lung and of metallic implants and especially for interventional MRI. It has already been shown that many of the devices used for interventional procedures (catheters, biopsy needles) can be used at lower fields without costly modifications. In addition, low-field MRI also allows the robust use of highly efficient sampling methods like spiral MRI. It is therefore safe to predict that low-field MRI is not only a cost-efficient compromise, but has the potential to open up new fields of application.