Login / Signup

Sensing Cd(II) Using a Disposable Optical Sensor Based on a Schiff Base Immobilisation on a Polymer-Inclusion Membrane. Applications in Water and Art Paint Samples.

Lorena Sánchez-PonceMaría Dolores Galindo-RiañoMaría José Casanueva-MarencoMaría Dolores Granado-CastroMargarita Díaz-de-Alba
Published in: Polymers (2021)
A disposable colour-changeable optical sensor based on an interesting polymer inclusion-membrane (PIM) was designed to determine Cd(II) ions in aqueous medium. The Schiff base 2-acetylpyridine benzoylhydrazone (2-APBH) immobilised on the polymer membrane was used as a sensing molecule. The amounts of the PIM components were optimised by a 3 2 fractional factorial design with two central points and two blocks. The best optical sensor composition consisted of 2.5 g of poly(vinylchloride) (PVC) as a base polymer, 3 mL of tributyl phosphate (TBP) as a plasticiser, and 0.02 g of 2-APBH as a reagent. The sensor showed a good linear response in the range from 0.02 mg L -1 (limit of detection) to 1 mg L -1 of Cd(II) under the following experimental conditions: pH 9.5 (adjusted using ammonium chloride buffer solution at 0.337 mol L -1 ), 60 min of exposure time plus 2 min of sonication (pulses at 2 s intervals), and 10 min of short-term stability. The relative standard deviation of the method was determined to be 4.04% for 0.4 mg L -1 of Cd(II). The optical sensor was successfully applied to the determination of Cd(II) in natural-water and art-paint samples.
Keyphrases
  • high resolution
  • high speed
  • nk cells
  • hiv infected
  • ionic liquid
  • liquid chromatography
  • neural network