High-impedance alerts with pulse generator-Lead mismatch.
Simon ChristieNada El TobgyColette M SeiferClarence KhooPublished in: Journal of arrhythmia (2022)
Use of mismatched leads with a Boston Scientific Accolade device was associated with an increased risk of undesirable changes in sensing polarity with occasional inappropriate pacing inhibition. Awareness of this interaction can allow for the institution of appropriate programming remedies and may increase scrutiny of the use of mismatched CIED systems.