Login / Signup

Evaluation of the interchangeability between the new fully-automated affinity chrome-mediated immunoassay (ACMIA) and the Quantitative Microsphere System (QMS) with a CE-IVD-certified LC-MS/MS assay for therapeutic drug monitoring of everolimus after solid organ transplantation.

Cristiano IalongoAnnamaria D'alessandroMaria SapioAntonio AngeloniOttavia Porzio
Published in: Clinical chemistry and laboratory medicine (2022)
ACMIA/EVRO but not QMS/EVER shows a lack of interchangeability with the CE-IVD-certified LC-MS/MS assay. We hypothesize that, as the ACMIA/EVRO uses an anti-sirolimus antibody, the under-corrected assigned value in the assay calibrators was not sufficient to reproduce the everolimus metabolites cross-reactivity occurring in real samples.
Keyphrases
  • high throughput
  • ms ms
  • deep learning
  • high resolution
  • single cell
  • energy transfer
  • cell therapy
  • mesenchymal stem cells
  • bone marrow