Login / Signup

Predicting ESI/MS Signal Change for Anions in Different Solvents.

Anneli KruveKarl Kaupmees
Published in: Analytical chemistry (2017)
LC/ESI/MS is a technique widely used for qualitative and quantitative analysis in various fields. However, quantification is currently possible only for compounds for which the standard substances are available, as the ionization efficiency of different compounds in ESI source differs by orders of magnitude. In this paper we present an approach for quantitative LC/ESI/MS analysis without standard substances. This approach relies on accurately predicting the ionization efficiencies in ESI source based on a model, which uses physicochemical parameters of analytes. Furthermore, the model has been made transferable between different mobile phases and instrument setups by using a suitable set of calibration compounds. This approach has been validated both in flow injection and chromatographic mode with gradient elution.
Keyphrases
  • ms ms
  • mass spectrometry
  • simultaneous determination
  • ionic liquid
  • drinking water
  • multiple sclerosis
  • high resolution
  • gas chromatography
  • data analysis