Optimizing calibration settings for accurate water equivalent path length assessment using flat panel proton radiography.

Carmen Seller OriaGabriel Guterres MarmittJeffrey FreeJohannes A LangendijkStefan BothAntje C KnopfArturs Meijers
Published in: Physics in medicine and biology (2021)
An optimal FP calibration in the framework of this study was established, characterized by 3 MeV ≤ ΔE ≤ 5 MeV and 2 mm ≤ ΔX ≤ 5 mm. Within these boundaries, highly accurate WEPL acquisitions using FP-PR are feasible and practical, holding the potential to assist future online range verification quality control procedures.