Validation of in vivo toenail measurements of manganese and mercury using a portable X-ray fluorescence device.
Aaron J SpechtXinxin ZhangAnna YoungVy T NguyenDavid C ChristianiDiana M CeballosJoseph G AllenJennifer WeuveLinda H NieMarc G WeisskopfPublished in: Journal of exposure science & environmental epidemiology (2021)
We identified that in vivo XRF metal measurements were able to discern exposure to manganese in welders and mercury in nail salon workers. We identified significant positive correlations between ICP-MS of clippings and in vivo XRF measures of both toenail manganese (R = 0.59, p = 0.02) and mercury (R = 0.74, p < 0.001), as well as between in vivo XRF toenail manganese and work history among the welders (R = 0.55, p = 0.03). We identified in vivo XRF detection limits to be 0.5 µg/g for mercury and 2.6 µg/g for manganese. Further work should elucidate differences in the timing of exposure using the in vivo XRF method over toenail clippings and modification of measurement time and x-ray setting to further decrease the detection limit. In vivo portable, XRF measurements can be used to effectively measure toenail Mn and Hg in occupational participants in real-time during study visits and at a fraction of the cost.