SEM Analysis and Micro-CT Evaluation of Four Dental Implants after Three Different Mechanical Requests-In Vitro Study.

Ana Sofia VinhasFilomena SalazarJosé Manuel MendesAntonio Sergio SilvaBlanca Ríos-CarrascoJosé Vicente Rios-SantosFrancisco Javier GilMariano Herrero-ClimentCarlos Manuel Aroso
Published in: Materials (Basel, Switzerland) (2024)
The application of mechanical compression loads caused deformation and contact marks in all models tested. Only internal connection implants revealed an axial displacement of the abutment into the implant, but at a general level, a clear intrusion of the abutment into the implant could only be confirmed in the Essential model, which obtained its maximal axial displacement with cyclic loading.