Validation of the eyeTelemed IOPvet indentation tonometer for use in dogs.

Lydia E KapellerAva G CabblePhillip N BuckmanChristine D HarmanAmanda L JacobsonFrank R LawrenceAndrás M Komáromy
Published in: Veterinary ophthalmology (2024)
The IOPvet is inaccurate in estimating canine IOP with a low sensitivity at identifying dogs with IOP > 30 mmHg. Canine-specific instrument revision is required to correctly identify elevated (yellow = 20-30 mmHg) and high (red >30 mmHg) IOPs.