Login / Signup

Back-calculation of keratometer index based on OCT data and raytracing - a Monte Carlo simulation.

Achim LangenbucherNóra SzentmáryJohannes WeisenseeAlan CaylessRupert MenapacePeter Hoffmann
Published in: Acta ophthalmologica (2021)
In the classical calculation, the keratometer index used for converting corneal radius to dioptric power uses several model assumptions. As these assumptions are not generally satisfied, corneal power cannot be calculated from corneal front surface radius alone. Considering all 6 input variables, the linear prediction model performs well and can be used if all input parameters are measured with a tomographer.
Keyphrases
  • monte carlo
  • optical coherence tomography
  • wound healing
  • cataract surgery
  • diabetic retinopathy
  • electronic health record
  • machine learning
  • artificial intelligence