The SAGA Light Source provides X-ray imaging resources based on high-intensity synchrotron radiation (SR) emitted from the superconducting wiggler at beamline 07 (BL07). By combining quasi-monochromatic SR obtained by the newly installed water-cooled metal filter and monochromatic SR selected by a Ge double-crystal monochromator (DCM) with high-resolution lens-coupled X-ray imagers, fast and low-dose micro-computed tomography (CT), fast phase-contrast CT using grating-based X-ray interferometry, and 2D micro-X-ray absorption fine structure analysis can be performed. In addition, by combining monochromatic SR obtained by a Si DCM with large-area fiber-coupled X-ray imagers, high-sensitivity phase-contrast CT using crystal-based X-ray interferometry can be performed. Low-temperature CT can be performed using the newly installed cryogenic system, and time-resolved analysis of the crystallinity of semiconductor devices in operation can be performed using a time-resolved topography system. The details of each instrument and imaging method, together with exemplary measurements, are presented.