Login / Signup

Association Between Intracochlear Electrode Design and Electrically-Evoked Compound Action Potential Measures in Cochlear Implant Users.

Jeong-Seo KimSung Hwa HongIl Joon Moon
Published in: Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery (2024)
Perimodiolar electrode groups yielded significantly narrower SOE half-widths and smaller CII than the lateral wall straight electrode group. This may indicate that the electrode array that hugged the modiolus had less overlap in neural excitation between adjacent electrodes, resulting in reduced channel interaction and potentially better spectral resolution than the electrode array positioned more laterally.
Keyphrases