Robotic-assisted total hip arthroplasty utilizing a fluoroscopy-guided system produced similar cup accuracy and precision relative to a computerized tomography-based robotic platform.

Christian B OngGraham B J BuchanChristian J Hecht IiArihiko KanajiDaniel O KendoffTarun K Jella
Published in: Journal of robotic surgery (2024)
Robotic assistance for total hip arthroplasty (THA) has been demonstrated to improve accuracy of acetabular cup placement relative to manual, unassisted technique. The purpose of this investigation was to compare the accuracy and precision between a fluoroscopy-based robotic total hip arthroplasty platform (FL-RTHA) and a computerized tomography-based (CT-RTHA) platform. The study included 98 consecutive FL-RTHA and 159 CT-RTHA procedures performed via direct anterior approach (DAA). All cases were performed for a pre-operative diagnosis of osteoarthritis, avascular necrosis, or rheumatoid arthritis. Primary outcome variables included cup implantation accuracy and precision (variance). Implantation accuracy was calculated as the absolute value of the difference between pre-operative target cup angles (inclination and anteversion) and the same post-operative angles. Percentage placement in the Lewinnek safe zone was also measured for both cohorts. The FL-RTHA and CT-RTHA cohorts demonstrated a 1.2° difference in absolute values for cup inclination accuracy (4.6° ± 3.6 vs. 3.4 ± 2.7; p = 0.005), and no difference in absolute values for cup anteversion accuracy (4.7° ± 4.1 vs. 4.6 ± 3.4; p = 0.991). Cohorts demonstrated similar precision for cup inclination and anteversion placement parameters, as well as equivalent Lewinnek safe zone placement. The use of a fluoroscopy-based robotic assistance platform for primary DAA THA resulted in similar accuracy and precision of acetabular cup placement when compared to a CT-based robotic assistance system.
Keyphrases
  • total hip arthroplasty
  • rheumatoid arthritis
  • computed tomography
  • minimally invasive
  • image quality
  • high throughput
  • contrast enhanced
  • dual energy
  • ultrasound guided
  • robot assisted
  • magnetic resonance imaging
  • pet ct