Validation of a Machine Learning Algorithm to Identify Pulmonary Vein Isolation during Ablation Procedures for the Treatment of Atrial Fibrillation: Results of the PVISION Study.
Jan De PooterLiesbeth TimmersSerge BovedaStephane CombesSebastian KnechtAlexandre AlmoradCarlos De AsmundisMattias DuytschaeverPublished in: Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology (2024)
This study validated an automated algorithm using machine learning (ML) to assess the isolation status of pulmonary veins in patients undergoing PVI with different ablation modalities. The algorithm reached an AUC of 92% with both sensitivity and specificity exceeding the primary study endpoints.