Developing a high-performance liquid chromatography fast and accurate method for quantification of silibinin.

Faezeh BakhshiOmmoleila MolaviMohammad Reza RashidiAli ShayanfarHassan Amini
Published in: BMC research notes (2019)
The retention time for silibinin was seen in 2.97 min after injection. The calibration curve was drawn and the established method demonstrated a linear ranged from 10 to 100 µg/ml, with a correlation coefficient of 0.996. The sensitivity of the developed method was 10 µg/ml. The accuracy calculated in the range of 88-105.9% and the precision (as relative standard deviation) was between 2.7 and 10.9%. These results demonstrate that the developed method can be a fast and accurate method for quantification of silibinin in aqueous samples.