Substrate and arrhythmia characterization using the multi-electrode Optrell mapping catheter for ventricular arrhythmia ablation-a single-center experience.
Jian Liang TanGustavo S GuandaliniMatthew C HymanJeffrey ArklesPasquale SantangeliRobert D SchallerFermin GarciaGregory SuppleDavid S FrankelSaman NazarianDavid LinDavid CallansFrancis E MarchlinskiTimothy M MarkmanPublished in: Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing (2023)
This novel MPMC can be used safely and effectively to create high-density maps in LV endocardium or epicardium. Limitations of the catheter include a longer wait time for matrix formation prior to starting point collection and challenges in manipulation in certain regions.