Refractive prediction of four different intraocular lens calculation formulas compared between new swept source optical coherence tomography and partial coherence interferometry.
Mi Yeon SongSung Rae NohKook Young KimPublished in: PloS one (2021)
SS-OCT and PCI demonstrated good agreement on biometric measurements; however, there were significant differences in some biometric values. These differences in some ocular biometrics can cause a difference in refractive error after cataract surgery. New type SS-OCT was not superior to the IOL power prediction calculated by PCI.