Investigating the effects of endogenous lipaemia on the measurement of sodium by indirect ion specific electrode potentiometry.

Eamon P McCarronEmma MurrayGareth C McKeemanStephen M CowardPaul HamiltonGrainne ConnollyBrona V Roberts
Published in: Annals of clinical biochemistry (2022)
) and it is clear that this is not comparable to endogenous lipaemia. To ensure patient safety, clinical laboratories should consider lowering the cut off for (L)-index that they use for reporting sodium concentration.