Differences between oscillometry measurements obtained by MostGraph-01 and MasterScreen-IOS in patients with asthma.

Sonoko HaradaNorihiro HaradaHitoshi SasanoYuki TanabeMasaki KotajimaYoshihiko SatoTomohito TakeshigeYoko KatsuraJun ItoRyo AtsutaHajime KurosawaKazuhisa Takahashi
Published in: PloS one (2024)
To our knowledge, this is the first report to compare MostGraph-01® and MasterScreen-IOS® devices using practical clinical data obtained in patients with asthma. The values obtained by both devices can be interpreted in a similar way, although there is slight variation. The conversion equations produced in this study may assist to compare the oscillometry measurements obtained by each of the two devices.