Comparability analysis of the HNT-1P Huvitz non contact tonometer for the measurement of intraocular pressure.
Alicia Cárceles MontoyaCarlos Enrique Monera LucasClaudia Patricia Tarazona JaimesDaniel Romero-ValeroAlejandro Moya MartínezJosé Juan Martínez ToldosPublished in: European journal of ophthalmology (2024)
HNT-1P (Huvitz) provides statistically significantly lower IOP values than those obtained with GAT. HNT-1P could be used for screening of ocular hypertension in postoperative patients. The IOP measurement obtained with HNT-1P should be confirmed with GAT. HNT-1P yields lower IOP values than those obtained with ICR.