Influences of oxidation ability on precision in nitrogen isotope measurements of organic reference materials using elemental analysis-isotope ratio mass spectrometry.

Nan WangJingyu LiuYang ZhangCuimei XiaDeju LinYangli CheJianzheng WuRui Bao
Published in: Rapid communications in mass spectrometry : RCM (2021)
The analytical conditions such as the oxygen flow rate and injection time or sample introduction time into the EA need to be adjusted depending on the combustion conditions of the RMs and samples to obtain a reliable accuracy of measurement. We recommend that when analyzing δ15 N of natural samples such as soil or marine sediment samples with low organic content, more oxygen should be injected to improve the combustion of the samples. In addition, CO2 should be removed from the sample gas stream before being introduced into the isotope ratio mass spectrometer when the EA is equipped with a single-quartz reactor.