Performance and application of heavy ion nuclear microbeam facility at the Nuclear Physics Institute in Řež, Czech Republic.

Oleksandr RomanenkoVladimir HavranekAnna MackovaMarie DavidkovaMariapompea CutroneoAlexander G PonomarevGyula NagyJames Stammers
Published in: The Review of scientific instruments (2019)
The Tandetron Laboratory of the Nuclear Physics Institute of the Czech Academy of Sciences is equipped with five beam lines associated with a 3 MV tandem electrostatic accelerator model 4130 MC from High Voltage Engineering Europa B.V. This accelerator is coupled with two duoplasmatron sources and a single sputter ion source and provides ions from hydrogen to gold. One of these lines is a nuclear microbeam facility, utilizing ion beams of micro- and sub-micro sizes for materials research by use of particle induced x-ray emission spectroscopy, particle induced gamma emission, Rutherford back-scattering spectroscopy, and scanning transmission ion microscopy methods as well as for ion beam writing. The major advantage of the presented microprobe is a possibility of 3D structure creation not only in polymer materials using light ions but also in other materials such as glass, ceramics, etc. by use of heavy ions. The focusing system allows focusing of charged particles with a maximum rigidity of 11 MeV amu/q2. The usual resolution in high and low current modes is 2 × 3 µm2 for a 100 pA and 0.3 × 0.5 µm2 for the 2000 ions/s of 2 MeV protons, respectively. A detailed facility description is given in the paper. The applications of focused beams of heavy ions as well as examples of light ions utilizing are also presented in the article.
Keyphrases