Dynamic Occlusion of Distal Ventriculoperitoneal Shunt Catheter after Infusion Port Placement: A New Shunt Malfunction.
Lacey M CarterCamille K MiltonKyle P O'ConnorArpan R ChakrabortyTressie M StephensChad A GlennPublished in: Journal of neurological surgery reports (2021)
Shunt failure requiring reintervention remains a common complication of hydrocephalus treatment. Here, we report a novel cause of mechanical shunt obstruction in an adult patient: position-dependent intermittent occlusion via an infusion port catheter. A 51-year-old woman with a grade II oligodendroglioma presented in a delayed fashion following surgery with a pseudomeningocele. She underwent ventriculoperitoneal shunt placement due to communicating hydrocephalus, resolving her pseudomeningocele. Shortly thereafter, she underwent placement of a subclavian infusion port at an outside institution. Her pseudomeningocele returned. Imaging demonstrated close proximity of her port catheter to the shunt catheter overlying the clavicle. Her shunt was tapped demonstrating a patent ventricular catheter with normal pressure. She underwent shunt exploration after her pseudomeningocele did not respond to valve adjustment. Intraoperative manometry demonstrated head position-dependent distal catheter obstruction. Repeat manometry following distal catheter revision demonstrated normal runoff independent of position. Her pseudomeningocele was resolved on follow-up. To our knowledge, this is the only reported case of intermittent, position-dependent distal catheter obstruction. Shunted patients with concern for malfunction following subclavian infusion port placement should be evaluated for possible dynamic obstruction of their distal catheter when the two catheters are in close proximity along the clavicle.