Apheresis machines variably overestimate mononuclear cell collection volume.
Dries DeerenMarleen NeyrinckHans VrielinkPublished in: Journal of clinical apheresis (2020)
Both Spectra Optia machines overestimated the collection volume after peripheral blood stem cell collection. Although the mean variation falls within the expected range, in some cases, this overestimation may be clinically relevant if no other method of measurement is used.