Experimental and Theoretical Study on the False Positive of Monomethyltin Determination in Toys Based on Gastric Juice Migration.
Lina HuangHanke LiYuhang LiPeng RenLezhou YiYang ChenTong-Mei MaPublished in: International journal of analytical chemistry (2020)
The phenomena of MMT false positive which occurred in the laboratory testing of toys is mainly because that TEMT is highly likely to be obtained from the reaction of inorganic tin and trace level of methylation agent impurities contained in the derivatization reagent-NaB(Et)4. To avoid MMT false positive, it is concluded that the maximum acceptable mole ratio of methylation agent impurities contained in NaB(Et)4 is approximately 0.028%. This research is helpful to be aware of methylation impurities and is favorable to avoid false judgment caused by MMT false positive in routine analysis of toys.