Login / Signup

Polyethylene wear of dual mobility cups: a comparative analysis based on patient-specific finite element modeling.

Julien WegrzynAlexander AntoniadisEhsan SarshariMatthieu BoubatAlexandre Terrier
Published in: International orthopaedics (2022)
When using XLPE instead of UHMWPE, dual mobility cup with a 22.2-mm-diameter femoral head produced a similar amount of volumetric wear than conventional acetabular component with a 32-mm-diameter femoral head against XLPE. Therefore, XLPE is advocated in dual mobility cup to improve its wear performance.
Keyphrases
  • finite element
  • total hip arthroplasty
  • optic nerve