Incorporating a Ferrous Polymer Target into Elastomeric Liners for Socket Fit Sensing in Prosthesis Users.
Ryan V CarterBrian G LarsenJake B McLeanJoseph L GarbiniJoan E SandersPublished in: Sensors (Basel, Switzerland) (2020)
Liner-to-socket distance measurement using inductive sensing may be an effective means to continuously monitor socket fit in people using trans-tibial prostheses. A practical limitation, however, is a means to incorporate a thin uniform-thickness layer of conductive or magnetically permeable target material into the wide range of prosthetic liner products that people with limb amputation commonly use. In this paper, a method is presented whereby a 0.50-mm thickness ferrous polymer made from a SEEPS polymer and iron powder that is formed adjacent to a 0.25-mm thick non-ferrous layer of SEEPS polymer is assembled between two sheets of elastic fabric material. Bench testing showed that the fabrication procedure achieved a root-mean-square error in the thickness of this construct of 58 μm, helping to create a consistent calibration result over the entire surface. The original fabric backing of an off-the-shelf prosthetic liner was removed and replaced with the developed construct. When worn in the shoe of an able-bodied participant for 7.5 h per day for 28 days, the sensor well maintained the shape of its calibration curve at the start of wear, but a distance offset (shifting of the y-intercept) was introduced that increased during the initial approximately 12 days of wear. When the distance offset was corrected, for the primary distance range of clinical interest for this application (0.00-5.00 mm), the sensor maintained its calibration within 4.4%. Before being used in clinical application for liner-to-socket distance monitoring, new ferrous liners may need to be pre-worn so as to achieve a consistent distance reference.