Accurate positioning of the Nellix EVAS system in the aortic neck may be challenging. Despite considerable experience with the system, still around half of the potential seal in the aortic neck was missed in the current series, without improvement over time. This should be considered during preoperative planning and may be a cause of a higher than expected complication rate. Detailed post-EVAS nonapposition surface can be determined with the described novel methodology that takes into account the sometimes irregularly shaped top of the sealing endobags.