The microbiological and physical properties of catheters for intermittent catheterization: a systematic review on the impact of reuse and cleaning.
Mark GrasdalMatthias WalterAndrei V KrassioukovPublished in: Spinal cord (2022)
Numerous cleaning methods resulted in the destruction of catheters. However, there are two reported cleaning methods, submersion for 5 min in 70% alcohol and the "Milton method", that eliminate bacterial colonization while leaving the physical properties of the catheters unchanged. While these cleaning methods are promising, each was published in just one study, therefore higher-powered / longitudinal studies confirming the safety and efficacy of these cleaning methods must be obtained before current clinical recommendations can be modified.