Electromagnetic interference from left ventricular assist devices detected in patients with implantable cardioverter-defibrillators.
Seth H SheldonMohammad-Ali JazayeriMichael PierpolineMoghniuddin MohammedValay ParikhAlexander RobinsonAmit NoheriaNicholas HaglundAndrew J SauerY Madhu ReddyPublished in: Journal of cardiovascular electrophysiology (2021)
ICD oversensing of EMI from LVADs is infrequent and can be managed with reprogramming the sensitivity. There was no significant difference in the R-wave SNR with IB versus DB ICD leads.